Generator R&S
UP300/350
E-1147.2759.00
6-120
Operating Manual, 11/2007
Selecting the Amplitude Distribution Function
Use
You can select the following amplitude distribution functions for the noise
signal:
Gaussian
Gaussian distribution functions (Gaussian factor = 5),
crest factor = 3.873
Rectangular
Rectangular distribution functions, most favourable ratio between the
RMS and peak value (crest factor = 1.732)
Triangular
Triangular distribution functions, crest factor = 2.450
Note:
The setting is always valid for both channels (Ch 1&2).
Selecting the
amplitude
distribution function
1.
Press the
function key
in the
menu.
A selection field containing the available settings is displayed. The
default setting is “Rectangular“.
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored and displayed in the parameter field.
Entering the Signal Amplitude
Use
You can enter the amplitude of the output signal as an RMS value. Here, the
maximum value depends on the amplitude distribution functions.
Selecting
the channel
1.
Press the numeric keys
1
,
2
, or
3
to select channel
Ch 1
,
Ch 2
, or both
channels
Ch 1&2
.
The selected channel is displayed in green in the channel display.
Entering the
signal amplitude
2.
Press the
function key
in the
menu.
An entry field containing the currently applicable setting is displayed. The
default setting is “100 mV (0.1 FS)“. At the same time, the function keys
are assigned various units of measurement.
Ch 1&2