![Rohde & Schwarz Smart Instruments UP300 Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/rohde-and-schwarz/smart-instruments-up300/smart-instruments-up300_operating-manual_1477754013.webp)
R&S UP300/350
Data Sheet
Operating Manual, 11/2007
1-13
E-1147.2759.00
DC voltage
Voltage range
0 V to ±33 V
Error limits
± (1 % of measured value +
0.5 % of measurement range)
Total harmonic distortion (THD)
Fundamental
20 Hz to 20 kHz
Frequency tuning
fi
xed through entered value,
auto-tuning to input signal
Weighted harmonics
up to 80 kHz
any combination of d2 to d9
Error limits
harmonics < 50 kHz
harmonics < 80 kHz
±0.7 dB
±1 dB
Inherent distortion
fundamental 1 kHz
fundamental 20 Hz to 5 kHz
fundamental 5 kHz to 15 kHz
fundamental 15 kHz to 20 kHz
< -100 dB
< -90 dB
< -85 dB
< -80 dB
Spectrum
bar graph showing signal and distortion
THD+N and SINAD
Fundamental
20 Hz to 20 kHz
Frequency tuning
fi
xed through entered value,
auto-tuning to input signal
Bandwidth weighting
fi
lters and sets of prede
fi
ned octave and third-octave
fi
lters; up to 3
fi
lters can be combined
Error limits
bandwidth < 22 kHz
bandwidth < 80 kHz
±0.8 dB
±1.4 dB
Inherent distortion
bandwidth 20 Hz to 22 kHz,
fundamental 1 kHz
bandwidth 20 Hz to 22 kHz,
fundamental 20 Hz to 5 kHz
bandwidth 20 Hz to 80 kHz,
fundamental 20 Hz to 20 kHz
< -95 dB + 4 µV
< -90 dB + 4 µV
< 80 dB + 8 µV
Spectrum post-FFT
of
fi
ltered signal
Difference frequency distortion
(DFD)
Measurement method
in accordance with IEC 268-3
or IEC 118
Frequency range
difference frequency
center frequency
80 Hz to 2 kHz
200 Hz to 80 kHz
Error limits
f
center
< 20 kHz
±0.5 dB
Inherent distortion
DFD d2, f
center
< 20 kHz
DFD d3, 5 kHz < f
center
< 20 kHz
< -105 dB
< -90 dB