Generator R&S
UP300/350
E-1147.2759.00
6-180
Operating Manual, 11/2007
The second
possibility:
Entering the step
size
3.
Press the
function key
in the
submenu
.
In case of linear spacing, the default setting is “200 Hz“. At the same
time, the function keys are assigned various units of measurement.
In case of logarithmic spacing, factor for step size is used.
4.
Enter a new value (
Ê
The permissible entry range depends on the frequency range; with
maximum frequency range it is:
Linear spacing:
(START-STOP)/1023
≤
STEP SIZE
≤
f
max
Logarithmic
spacing
(
STOP/START
)
/1023
≤
STEP SIZE
≤
STOP/START
The new setting is stored and displayed in the parameter field.