R&S FSH
Spectrum Analyzer Mode
Performing Spectrum Measurements
Operating Manual 1173.6275.12 - 30
82
► Press the MEAS key.
► Press the "Meas Mode" softkey.
► Select the "Spectrogram" menu item.
The R&S FSH starts the spectrogram result display.
By default, the spectrogram result display consists of two windows. The upper window
shows the measured spectrum as a trace line. The lower window shows the
measurement results in a spectrogram. The chronological information in the
spectrogram is restricted by the internal memory of the R&S FSH. The R&S FSH
stores 1024 frames or spectrums that have been measured in its memory. As the
height of the display is smaller, some of the data becomes invisible after a time.
3.1.10.1 Controlling the Spectrogram Update
The spectrogram starts running as soon as you enter the spectrogram mode, if you are
in continuous sweep mode. If you are in single sweep mode, the R&S FSH does not
add a line to the spectrogram until you initiate the next single sweep.
You can stop the update of the spectrogram in continuous sweep mode.
► Press the MEAS key.
► Press the "Hold" softkey.
Note that in continuous sweep mode the measurement in the spectrum result
display does not stop. The trace in the upper window still updates continuously.
Only the spectrogram view stops.
► Press the "Hold" softkey again.
The R&S FSH resumes updating the spectrogram.
The spectrogram result display is filled with results until you change a measurement
setting. As soon as a setting is changed, the spectrogram clears and starts to fill again.
You can also clear the spectrogram manually.
► Press the MEAS key.
► Press the "Clear" softkey.
3.1.10.2 Browsing Through the Signal History
There are two ways to view parts of the measurement result history that have moved
outside the visible area of the spectrogram.
► Press the MEAS key.
► Press the "Spectrogram Settings" softkey.
► Select the "[ ] Spectrogram Full Screen" menu item.
The R&S FSH now uses the full diagram area of the screen for the spectrogram.
The number of lines in the spectrogram and therefore the displayed time period
more than doubles.