R&S FSH
Network Analyzer Mode
Performing Vector Measurements (R&S FSH-K42)
Operating Manual 1173.6275.12 - 30
192
6.4.1 Measuring the Transmission
The example included here is based on a transmission measurement of a two-port
filter. The filter works in the frequency from of 1920 MHz to 1980 MHz.
Test setup
► Connect the input of the DUT to the tracking generator output (port 2).
► Connect the output of the DUT to the RF input (port 1).
Preset the R&S FSH
Before starting the measurement procedure, preset the R&S FSH to restore the default
configuration and connect the filter between the measurement ports.
► Press the PRESET key.
► Connect the DUT.
Start vector measurements.
► Press the MEAS key.
► Press the "Meas Mode" softkey.
► Select the "Vector" menu item.
Select the type of transmission measurement
► Press the "Result Display" softkey.
► Select the "Transmission Fwd (Port 1
2)" menu item for transmission
measurements in forward direction or "Transmission Rev (Port 1
2)" menu item
for transmission measurements in reverse direction.
Simultaneous display of several results
You can also display two results simultaneously in a single diagram. In that case, the
R&S FSH shows two traces with a different color in the result display.
► Press the MEAS key.
► Press the "Result Display” softkey.
► Select one of the result combinations to be displayed in a single diagram.
-
Reflection S11 + S22
-
Transmission S21 + S12
-
Transmission / Reflection S21 + S11
-
Transmission / Reflection S12 + S22
The information, which trace carries which result, is indicated by a label in the
diagram area.
Note that the simultaneous display of more than one result in a single diagram is
only possible with a high accuracy calibration.