R&S FSH
Receiver Mode (R&S FSH-K43)
Selecting the Measurement Mode
Operating Manual 1173.6275.12 - 30
233
8.1.1.1 Defining the Receive Frequency
In single frequency measurement mode, the R&S FSH determines the power level of a
single frequency only. The available frequency range depends on the R&S FSH model
you are using.
► Press the FREQ key.
► Press the "Freq" softkey.
Enter the frequency you want to measure.
For a quick change of frequencies with the rotary knob, define a frequency stepsize.
► Press the "Freq Stepsize" softkey.
Enter a frequency stepsize.
For measurements on systems that use channels instead of single frequencies, you
can also load a channel table and measure the channel power.
► Press the "Channel Mode" softkey.
The R&S FSH opens a dialog box to select a channel table.
For measurements based upon a channel table, you can select a single channel
instead of a single receive frequency.
► Press the "Channel" softkey.
► Enter a channel number.
8.1.1.2 Customizing the Bargraph Aspects
The bargraph is a graphical representation of the power level with one dimension, the
power level of the receive freuqency. The R&S FSH provides several ways to
customize the aspects of the bargraph.
Selecting the Unit
By default, the unit the R&S FSH uses in receiver mode for the measured power levels
in general is dBµV. In addition to dBµV, the receiver mode provides other units as well.
► Press the AMPT key.
► Press the "Unit" softkey.
Select the unit you need.