![Rohde & Schwarz FSH13 Скачать руководство пользователя страница 150](http://html.mh-extra.com/html/rohde-and-schwarz/fsh13/fsh13_operating-manual_1477832150.webp)
R&S FSH
Power Meter
Using a Power Sensor
Operating Manual 1173.6275.12 - 30
135
4.1.1 Connecting a Power Sensor
The R&S FSH controls and powers the power sensors via a special interface on the
top of the instrument. You can also connect a power sensor via the USB interface on
the right.
If you are using the R&S FSH-Z1 and -Z18 power sensors, connect the power sensor
cable to the power sensor interface and screw it into position. Power sensors of the
R&S NRP product line are controlled via the USB interface with a passive USB
adapter.
After connecting the power sensor to the R&S FSH, you can connect the DUT to the N-
connector of the power sensor.
Risk of damage to the power sensor
Before you start to work with the power sensor, make sure that the continuous power
applied to the input of the power sensor does not exceed a certain level.
Refer to the documentation of the power sensor for more information on the maximum
input power.
If the R&S FSH recognizes a power sensor, it sets up a connection via the interface
and after a few seconds shows the measured power. It displays the type of the power
sensor in the display header.
If no power sensor has been connected or it is not connected appropriately, the
R&S FSH shows nothing.
If there are communication problems between the R&S FSH and the power sensor, the
R&S FSH displays one of the following error messages that indicate the possible
cause.
Message
Cause
Remedy
Error in zeroing: signal at sensor
A signal was present at the power sensor
when zeroing was performed.
Unscrew the power sensor from the device
under test and repeat zeroing.
Warning: Input overloaded
The power at the input of the power sensor
exceeds the permitted power (23 dBm =
200 mW).
Reduce the power at the sensor input.
Power sensor hardware error
Communication error between the R&S FSH
and the power sensor.
Unscrew the sensor from the R&S FSH and
check the connectors.
If the problem persists, contact a Rohde &
Schwarz service center.
Power sensor error
The power sensor signals an error to the
R&S FSH.
Contact a Rohde & Schwarz service center.
Unknown power sensor model
connected
TheR&S FSH cannot identify the device
connected to the power sensor interface.