Network Analyzer Mode
Calibrating Measurements
Operating Manual 1173.6275.12 - 30
182
You can restore the factory calibration manually any time.
-
Press the "Calibration" softkey.
-
Select the "User Calibration Off" menu item.
● (fcal?)
The R&S FSH uses factory calibration. However, the calibration is not accurate
because the power of the tracking generator and the attenuation at the RF input
are not in line with the default settings. In that case you should repeat calibration.
● (cal)
The R&S FSH uses user calibration. To get that state you have to perform either a
full 1-port or a full 2-port calibration.
● (cal?)
The R&S FSH uses user calibration. However, the calibration is not accurate
because the TG power and receiver attenuation are not in line with the settings at
the time it has been calibrated. In that case you should repeat calibration.
● (norm) (network analysis only)
The R&S FSH uses normalization. To get that state you have to normalize the
transmission.
● (norm?) (network analysis only)
The R&S FSH uses normalization. However, the normalization is not accurate
because the TG power and receiver attenuation are not in line with the settings at
the time it has been calibrated. In that case you should repeat calibration.
● (interp) (network analysis only)
The R&S FSH interpolates the correction data between the reference points of the
calibration. Interpolation is used when you change one of the frequency
parameters (start, stop or center frequency). In that case the distribution of the
measurement points is different to the distribution during calibration. This could
result in an increasing measurement uncertainty.
When calibration has become invalid for any reason or the calibration data has
changed, you can restore the most recent calibration that was valid.
► Press the "Calibration" softkey.
► Select the "Restore Calibration Settings" menu item.
The R&S FSH restores the calibration data and the frequency settings that were
active.