Distance-to-Fault Mode (R&S FSH-K41)
Configuring Cable and Antenna Tests
Operating Manual 1173.6275.12 - 30
223
7.2.1.3 Working with a DTF List
The DTF list shows the results of the distance to fault measurement in numerical form.
► Press the MEAS key.
► Press the "DTF Settings" softkey.
► Press the "DTF List" softkey.
The R&S FSH opens a table that lists the peaks it has found during the
measurement.
The table contains the following information.
● Peak
Shows the peak the results refer to.
● Distance
Shows the distance from the measurement plane to the peak.
● Return Loss
Shows the magnitude of the peak.
To limit the information in the list, you can define a threshold for the DTF list, so that
only peaks above a certain level are included in the list.
► Press the MEAS key.
► Press the "DTF Settings" softkey.
► Press the "DTF List Threshold" softkey.
The R&S FSH opens an input field to define a threshold in dB.
► Enter the threshold you want.
The R&S FSH now only shows the peaks that are above the level you have
entered.