Spectrum Analyzer Mode
Performing Spectrum Measurements
Operating Manual 1173.6275.12 - 30
79
Note that the frequency range of the actual measurement depends on the start and
stop frequency you have set for the carrier. Correct measurement results are therefore
only possible if the frequency ranges of the measurement are inside the current span
of the R&S FSH.
Markers in the Spurious Emission measurement
In addition to the normal marker functionality of the R&S FSH, the Spurious Emission
measurement provides special markers labeled P1 to Px.
The R&S FSH activates and positions these special markers automatically after
displaying the trace. Each of these markers marks the peak level in each Spurious
Emission range.
The numerical marker information (frequency and level) for these markers is part of the
result table.
Markers P1 to P4 are fix and always remain on the peak level of the corresponding
subrange. If you'd like to analyze other locations on the trace, use normal marker and
deltamarker functionality. For more information see "
Using Markers and Deltamarkers
on page 116.
3.1.9.1 Defining the Channel Bandwidth
Several frequency bands have been defined for 3GPP transmissions. Depending on
the frequency band you are testing, you have to adjust the start and stop frequencies
of the carriers.
► Press the MEAS key.
► Press the "Carrier Start" softkey.
The R&S FSH opens an input field to define the lower frequency of the channel.
► Press the "Carrier Stop" softkey.
The R&S FSH opens an input field to define the upper frequency of the channel.
Note that the standard defines spurious emissions as transmission outside the
frequency band 12.5 MHz below the first carrier and 12.5 MHz above the last carrier.
Keep these values in mind when defining the start and stop frequencies of the carrier.
3.1.9.2 Viewing the Results in a Table
You can add a table to the display that shows the measurement results in numerical
form.
► Press the MEAS key.
► Press the "View List" softkey.
The R&S FSH shows a list above
the trace diagram.