R&S FSH
Spectrum Analyzer Mode
Using Transducer Factors
Operating Manual 1173.6275.12 - 30
131
You can select two transducer factors, a primary transducer and a secondary
transducer. If a transducer factor is active, the menu item has an [X] in front of it.
► Select the "Select Primary Transducer" menu item.
The R&S FSH opens a dialog box to select the transducer factor.
► Select the transducer factor you need.
► Confirm the selection with the "Select" softkey.
The R&S FSH shows the name of the active transducer on the display.
An example would be the transducer factor of the R&S HL223 antenna that is defined
between 200 MHz and 1300 MHz. The R&S FSH therefore displays the noise in this
frequency range as a function of frequency incremented by the transducer factor.
Outside the transducer range, the R&S FSH sets the transducer factor at zero, i.e.
measurements in this range do not yield conclusive results.
You can select a second transducer factor with the "Select Secondary Transducer"
menu item. The secondary transducer factor in that case is added to the first. The unit
of the second transducer factor must always be the relative unit dB as otherwise an
addition would not be useful. When you select a secondary transducer factor, the
dialog box shows only those transducer factors that have dB as their unit.
3.4.1 Unit for Measurements with Transducers
If the unit of the transducer is dB, the units dBm, dBmV or dBµV remain unchanged.
The linear units Volt and Watt are not permissible. They are deactivated in the units
menu.
If the unit of the transducer is dBµV/m or dBµA/m, this unit is also used for the
R&S FSH level display. This means that both the level axis of the diagram and the
level at the marker position are assigned the unit of the transducer. If dBµV/m is
selected as the transducer unit, a switch to absolute level indication in V/m is possible.
► Press the AMPT key.
► Press the "Unit" softkey.
► Select the V(olt) menu item.
If you are using a transducer with the unit dBµA/m, it is not possible to select another
unit. Level indication is entirely in dBµA/m.
3.4.2 Setting the Reference Level
The transducer shifts the trace by its value as a function of frequency. Positive
transducer values increase the level, negative values reduce it. To ensure that the
trace is always within the diagram, the R&S FSH adjusts the reference level
accordingly. The reference level is shifted by the maximum transducer value in the
positive or negative direction.