R&S FSH
Network Analyzer Mode
Vector Voltmeter (R&S FSH-K45)
Operating Manual 1173.6275.12 - 30
211
6.6.2 Performing Measurements
With the vector voltmeter you can measure the reflection on port 1 and the
transmission in reverse direction.
The example included here is based on a transmission measurement of a two-port
filter. The filter works in the frequency from of 1920 MHz to 1980 MHz.
Select the type of reflection measurement
► Press the "Result Display" softkey.
► Select the "Reflection Port 1" menu item for reflection measurements on port 1 or
the "Transmission Rev (Port 2
1)" menu item for transmission measurements in
reverse direction.
Define the frequency parameters
Before calibrating the measurement, you should define the frequency parameters to
avoid interpolation of the results.
► Press the FREQ key.
The R&S FSH opens an input field to set the center frequency.
► Enter a frequency of 1950 MHz.
6.6.2.1 Performing Transmission Measurements
Calibrate the measurement for transmission measurements.
► Press the "Calibration" softkey.
► Select the "Transmission Rev
(Port 2
1)" menu item.
► Perform calibration. For more
information see "
".
► Reconnect the DUT.
The R&S FSH shows the
characteristics of the DUT
(magnitude and phase) as
numerical values.
You can change the measurement configuration (e.g. the sweep time) or format
(see "
Selecting the Measurement Format
") without affecting the accuracy of the
measurement.