![Rohde & Schwarz FSH13 Скачать руководство пользователя страница 133](http://html.mh-extra.com/html/rohde-and-schwarz/fsh13/fsh13_operating-manual_1477832133.webp)
R&S FSH
Spectrum Analyzer Mode
Configuring Spectrum Measurements
Operating Manual 1173.6275.12 - 30
118
► Confirm the delta marker position with the ENTER key.
The delta marker input field closes.
► To add more markers, press the "New Marker" softkey several times until you have
the number of markers you want in the display.
3.2.6.4 Selecting the Marker Type
When you add new markers, they will be delta markers by default. Their coordinates
are relative to the first marker (M1). You can turn delta markers into normal markers if
you need absolute information about the marker position.
► Press the MARKER key.
► Select the delta marker you want to convert with the "Select Marker" softkey.
The corresponding marker symbol turns red and the marker input field opens.
► Press the "Marker Type" softkey.
The delta marker turns into a normal marker. Its label changes accordingly (e.g. D2
to M2) and its coordinates are now absolute values.
3.2.6.5 Automatic Positioning of Markers
The R&S FSH offers functions that make setting the markers easier or allow to make
instrument settings on the basis of the current marker position:
● "Set to Peak"
The Peak function places the active marker or the delta marker on the highest level
value of the trace.
● "Set to Next Peak"
The Next Peak function places the active marker or delta marker on the next
highest level value of the trace, relative to its current position.
● "Set to Minimum"
The Minimum function places the active marker or delta marker on the lowest
value of the trace.
► Press the MKR
key.
► Press the "Set to Peak", "Set to Next Peak" or "Set to Minimum" softkey.
The R&S FSH positions the marker accordingly.