Network Analyzer Mode
Vector Voltmeter (R&S FSH-K45)
Operating Manual 1173.6275.12 - 30
212
6.6.2.2 Performing Reflection Measurements
Calibrate the measurement for reflection measurements.
► Press the "Calibration" softkey.
► Select the "Reflection Port 1" menu item.
► Perform calibration. For more information see "
".
► Reconnect the DUT.
The R&S FSH shows the characteristics of the DUT (magnitude and phase) as
numerical values.
You can change the measurement configuration (e.g. the sweep time) or format
(see "
Selecting the Measurement Format
") without affecting the accuracy of the
measurement.
Note that for calibration to remain valid, frequency parameters, bandwidth and
attenuation have to remain the same.