R&S FSH
Spectrum Analyzer Mode
Performing Spectrum Measurements
Operating Manual 1173.6275.12 - 30
73
3.1.6.3 Viewing the Results in a Table
You can add a table to the display that shows the measurement results in numerical
form.
► Press the MEAS key.
► Press the "View List" softkey.
The R&S FSH shows a list above
the trace diagram.
If the list contains more than four entries, you can scroll through the list with the
rotary knob or the cursor keys to see the other results. Note that scrolling works
only if there is no active input field.
The list contains the following information:
● Tx Power
Power level of the transmission channel.
● Tx Bandwidth
Bandwidth of the transmission channel.
● PASS / FAIL information
If the signal is within the limits of the spectral mask, the R&S FSH shows
,
if not it shows
.
● Range [Hz]
Frequency range. The first number is the start frequency, the second number the
stop frequency of each defined frequency range. The character following the
number indicates the unit (k = kHz, M = MHz, G = GHz)
● RBW [Hz]
Resolution bandwidth the corresponding frequency range is measured with.
● Peak
Number of the peak marker (Px).
● Freq [Hz]
Frequency of the peak level that has been measured in each frequency range
● Power Abs
Absolute peak power in the corresponding frequency range.
● Power Rel
Relative peak power in relation to the channel power of the reference channel.
● Δ Limit
Minimum distance from the limit line to the trace in the corresponding frequency
range. Negative values or a zero indicate a passed SEM limit check, positive
values indicate a violation of the limit check.