Calibrating Measurements
Operating Manual 1173.6275.12 - 30
186
► Press the MEAS key.
► Press the "Calibration" softkey.
► Select the "Select Calkit [ ]" menu item.
The R&S FSH opens a dialog box to select the calibration kit. The list of supported
calibration kits covers those supported by the R&S FSH and those that you have
customized.
► Select the calibration kit you are using.
The R&S FSH uses the characteristics specified for the calibration kit.
It also shows the calibration kit currently in use next to the "Select Calkit [ ]" menu
item.
You can also use calibration kits other than the R&S FSH-Z28 or -Z29 in combination
with their characteristics. However, make sure that the electrical length of the
calibration kit in use is similar to that of the R&S FSH-Z28 or -Z29. The electrical length
of the "Open" and "Short" of the R&S FSH calibration kits is 5.27 mm. If the electrical
length is different, it may cause additional phase error.
Some measurement setups may include additional cables or adapters that may have
an additional electrical length. To correct their phase error, you can include the
electrical length of additional test equipment. The R&S FSH distinguishes if the
equipment is connected to port 1 or port 2.
► Press the MEAS key.
► Press the "Calibration" softkey.
► Select the "Offset Length Port 1" or "Offset Length Port 2" menu item.
The R&S FSH opens an input field to define the electrical length of the accessory.
► Enter the electrical length.
The R&S FSH now includes the electrical length of the accessory for phase
measurements.