R&S FSH
Digital Modulation Analyzer
Measurements on TD-SCDMA Signals
Operating Manual 1173.6275.12 - 30
302
9.7.5 The Time Domain Power Result Display
If you equip the R&S FSH with option R&S FSH-K48E, the Time Domain Power result
display becomes available.
The Time Domain Power result display shows the power of the signal over one TD-
SCDMA subframe (5.42 ms or 7 time slots, including the pilot time slots). This
measurement is useful to determine which time slots contain power.
Screen layout of the time domain power result display
1
Current type of result display
2
Currently selected standard
3
Header table
4
Result table
5
Diagram
6
TD-SCDMA softkey menu
The result display consists of a result table in the upper part of the display and a
diagram in the lower half of the display.
The diagram contains a trace that shows the power of the signal over one TD-SCDMA
subframe. The time slots in the frame are represented by blue vertical lines. Each time
slot is also labeled with a number. A subframe consists of 7 normal time slots,
therefore the range of numbers is 0 through 6. The trace always begins with time slot
0.
Note that the diagram also shows three time slots which are not numbered and have a
shorter duration than the normal time slots. These are the special time slots that
belong to each subframe: two time slots that conatin the pilot information and one time
slot that serves as a guard period.