R&S FSH
Working with the Measurement Wizard
Using the Measurement Wizard
Operating Manual 1173.6275.12 - 30
45
● Site Name
Location of the measurement. This field is available on the R&S FSH only.
● Comments
Comments about the measurement, e.g. the external conditions during the
measurement.
● GPS Position
Shows the GPS position, if you have connected a GPS receiver. Pressing the
ENTER key on this field results in an update of the GPS coordinates.
● Use Wizard Cable Settings
Determines if you want to use the cable characteristics as defined in the
measurement set or if you want to be able change cable characteristics on site.
Select "Yes" to use the predefined cable characteristics. In That case the
parameters below will be locked.
● Cable Model
Cable model that you perform the measurement on. You can define a cable model
with the R&S FSH4View software, but can also change the cable model on short
notice, if necessary.
● Clear Cable Model
Deactivates the currently active cable model.
● Cable Length
Length of the cable that you perform the measurement on.
● Calibration
Calibration method to use before the measurement starts. This is a read only field,
the calibration method has to be defined with R&S FSH4View.
You have to calibrate the R&S FSH before you can begin with the measurement
sequence defined in the wizard. If the R&S FSH has already been calibrated with
the defined routine prior to starting the wizard, the R&S FSH skips the calibration
and directly starts the measurement.
● Measurements
List of all individual measurements (datasets) that need to be performed for
successful completion of the measurement task. The list also shows the
measurements that still need to be performed.
Some parameters of the measurement setup you can still change directly on the
R&S FSH. These are mainly parameters whose details may not be available when you
define the measurement set or whose details may differ depending on the
measurement site, e.g. the cable length or the cable model if it is different to the one
defined previously.
► To change a parameter, select it with the cursor keys and activate the
corresponding input field with the ENTER key.
► Update all parameters that are not correct for the current measurement.