R&S FSH
Spectrum Analyzer Mode
Performing Spectrum Measurements
Operating Manual 1173.6275.12 - 30
93
If you are using the extension cable R&S TS-EMFZ2, you have to take this into
account as a secondary transducer.
► Press the "Transducer" softkey.
► Select the "Select Secondary Transducer" menu item.
The R&S FSH opens a dialog box to select transducer factors with the unit dB.
► Select the transducer factor for the extension cable.
► Confirm the selection with the "Select" softkey.
For more information see "
You can create and edit transducer factor with the R&S FSH4View software package
and then transfer them into the internal memory of the R&S FSH. Each transducer
factor may consist of up to 1000 reference values.
Display of the antenna directions
An isotropic antenna consists of three orthogonal elements. Each of these elements
measures the field strength from a different direction (x-, y- and z-axis).
Decrease of measurement speed
Because the R&S FSH performs a measurement on each of the three antenna axes,
the update rate of the results decreases slightly.
You can select to display different aspects of the measurement.
► Press the "Meas Mode" softkey.
► Select the "Iso Direction [ ]" menu item.
The R&S FSH opens a new menu to select the measurement aspect.
● "Auto"
Shows the total field strength over all three antenna axes.
The displayed result is a combination of the results for each antenna element. After
measuring each of the three directions individually, the R&S FSH calculates the
total field strength (E
r
) based on the results for each antenna element.
2
2
2
z
y
x
r
E
E
E
E
● "X"
Shows the field strength measured on the antenna's x-axis only.
● "Y"
Shows the field strength measured on the antenna's y-axis only.
● "Z"
Shows the field strength measured on the antenna's z-axis only.