148 • Using Maintenance Mode
Gas Tracer Operator’s Manual, CSA Version
Updating the Calibration Reminder Setting
With CL RMNDR set to ON (factory setting), the Gas Tracer will give an
indication at start up if it is due for calibration. The type of indication will depend
on the CL EXPRD setting (see below).
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the CL RMNDR menu item.
Press and release the POWER ENTER button. The screen will indicate the
current setting.
2. Use the AIR
▲
or (SHIFT)
▼
button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and return to
the main menu.
Updating the Calibration Expired Action Setting
This item defines what indication is given during start up when calibration is due
and CL RMNDR is set to ON.
With CL EXPRD set to CONFIRM (factory setting), the Gas Tracer will give an
indication at start up if calibration is due and require the user to press the RESET
SILENCE button to continue.
With CL EXPRD set to NOT USE, if the unit is due for calibration the Gas Tracer
will give an indication at start up that calibration is due and the unit cannot be
used until it is calibrated. With this setting, it is necessary to calibrate the
instrument from Maintenance Mode if calibration is not performed before it is
due. See “Calibrating Using Auto Calibration” on page 115, “Calibrating Using
Single Calibration” on page 118, or “Calibrating and Bump Testing the PPM
Combustible Sensor” on page 125 for instructions.
With CL EXPRD set to NO EFFECT, no confirmation during startup or calibration
is required to use the Gas Tracer if calibration is overdue.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the CL EXPRD menu item.
Press and release the POWER ENTER button. The screen will indicate the
current setting.
On
MAINTENANCE
CL RMNDR
CAL
SEL
MAINTENANCE
CONFIRM