Gas Tracer Operator’s Manual, CSA Version
Display Mode • 55
User ID Screen
This screen always appears in Display Mode, regardless of the ID DISP setting in
the Maintenance Mode menu. Each user ID can contain up to 16 characters but
only the first 8 of those characters are visible in the instrument. Be sure that all
user IDs are distinguishable from each other by viewing only the first 8
characters. For more information, see the GX-2012 Data Logger Management
Program manual. Uppercase letters, numbers, asterisks (*), dashes (-), and a blank
space are available characters. The Gas Tracer can store 128 user IDs. The bottom
line of this screen alternates between YES/ENT., the current user ID, and NO /
DISP.
Use this screen to identify or change the user. The user ID provides a way to
identify the user during a data logging session. The user ID is saved to the data
logger for that session when you turn off the Gas Tracer or change users during
operation. This allows you to change the user ID during operation and have each
user ID that was used during an operating session saved for the corresponding
data.
NOTE:
As shipped from the factory, the user ID on this screen is shown as
-- -- -- -- -- -- until you enter this screen and select a user ID.
To change a user:
1. After entering Display Mode and proceeding to the User ID screen, press and
release the POWER ENTER button. The bottom line of this screen alternates
between YES/ENT, the current user ID, and NO /DISP.
2. Use the AIR
▲
and (SHIFT)
▼
buttons to scroll through the 128 possible
user IDs.
3. Press the POWER ENTER button to select a user ID.
U Id
YES/ENT.
SEL
U Id
USER--001
SEL