Gas Tracer Operator’s Manual, CSA Version
Using Maintenance Mode • 155
Updating the Bar Hole Sampling Time
The Bar Hole Sampling Time is the amount of time that the Gas Tracer samples a
bar hole when being used in Bar Hole Mode. The 3 selections are 30, 45, and 60
seconds. The factory setting is 30 seconds.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the B.H. TIME menu item.
2. Press and release the POWER ENTER button. The screen will indicate the
current setting.
3. Use the AIR
▲
or (SHIFT)
▼
button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and return to
the main menu.
Updating the Leak Check Mode CO Channel Display Setting
With CO DISP set to OFF (factory setting), the CO channel is not displayed in
Leak Check Mode but is still displayed in Normal Mode.
With CO DISP set to On, the CO channel will be displayed in Leak Check Mode
but there will be no CO alarms and the CO channel reading will not be saved to
any snap log data taken during the operating session.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the CO DISP menu item.
MAINTENANCE
B.H. TIME
30
MAINTENANCE
B.H. TIME
MAINTENANCE
CO DISP