84 • Calibration Mode
Gas Tracer Operator’s Manual, CSA Version
9. When AUTO CAL in the BUMP--SET menu item in Maintenance Mode is set
to ON:
If the ppm combustible sensor passes the bump test,
•
The following screen appears:
•
To view the bump test gas reading press the AIR
▲
or (SHIFT)
▼
button.
You can scroll between the bump test results and the bump test gas
reading with the AIR
▲
or (SHIFT)
▼
button.
•
To return to the BUMP screen in PPM CAL, press POWER ENTER.
•
Disconnect the humidifier from the probe.
NOTE:
Grasp the humidifier tube assembly by the clear tubing at the end, not
the humidifier tube itself.
P
ppm
MAINTENANCE
CH
4
BUMP
500
ppm
MAINTENANCE
CH
4
BUMP
MAINTENANCE
BUMP