60 • Display Mode
Gas Tracer Operator’s Manual, CSA Version
The gas readings that were taken during the snap log are displayed and the
Station ID that was in use during the snap log is displayed along the bottom of
the screen. Snap log data that was taken in Leak Check Mode operation will
have 2 sequential screens that are part of one data set. One screen is for the
base reading and the other screen is for the peak reading. The peak reading
value will have PEAK displayed below the peak bar. If a base data point
appears without a peak data point after it, it is because a peak data point was
not taken. You can use the AIR
▲
and (SHIFT)
▼
buttons to scroll through the
different snap logs. The gas readings and station ID will change but the snap
log ID is not visible from this screen.
Peak Bar Select Screen
The Peak Bar Select Screen allows you to turn the peak bar on or off. If the
function is turned on, the peak bar appears along the left side of the screen during
Normal Operation and on the Peak Screen in Display Mode. It shows the peak
readings for each channel in bar graph format. The bottom line of this screen
alternates between the current setting, NO /DISP, and YES/ENT.
To adjust the peak bar selection:
1. Use the DISPLAY (ADJ) button to scroll to the Peak Bar Select Screen. The
screen indicates the current setting.
2. Press and release the POWER ENTER button.
bAr
SEL
BAR OFF
bAr
SEL
BAR ON
On
PEAK BAR