110 • Leak Testing
Gas Tracer Operator’s Manual, CSA Version
The gas readings that were taken during the snap log are displayed and the
Station ID that was in use during the snap log is displayed along the bottom of
the screen. Snap log data that was taken in Leak Check Mode operation will
have 2 sequential screens that are part of one data set. One screen is for the
base reading and the other screen is for the peak reading. The peak reading
value will have PEAK displayed below the peak bar. If a base data point
appears without a peak data point after it, it is because a peak data point was
not taken. You can use the AIR
▲
and (SHIFT)
▼
buttons to scroll through the
different snap logs. The gas readings and station ID will change but the snap
log ID is not visible from this screen.
6. To exit snap log data viewing and return to Leak Check Mode Operation,
press and release the DISPLAY (ADJ) button.
Turning Off the Gas Tracer, Leak Check Mode
1. Press and hold the POWER ENTER button.
2. The buzzer will pulse for about three seconds and TURN OFF will appear at
the bottom of the screen.
3. Release the button when TURN OFF disappears.