Gas Tracer Operator’s Manual, CSA Version
Using Maintenance Mode • 149
2. Use the AIR
▲
or (SHIFT)
▼
button to display the desired setting along the
bottom of the screen.
3. Press and release the POWER ENTER button to save the setting and return to
the main menu.
Updating the Calibration Check Setting
This item setting defines which sensors are used to update the calibration
reminder screen. C--CHECK can be set to 4-GAS or ALL. If C--CHECK is set to 4-
GAS (factory setting), the calibration dates for only the standard sensors (LEL, O
2
,
and CO) are used to determine if calibration is due. If C--CHECK is set to ALL,
the calibration dates for the standard sensors along with the %volume sensor and
the ppm combustible gas sensor are used to determine if a calibration is due.
1. Use the AIR
or (SHIFT)
buttons to display the C--CHECK menu item.
Press and release the POWER ENTER button. The screen will indicate the
current setting.
2. Use the AIR
or (SHIFT)
button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and return to
the main menu.
Updating the Bump Test Interval
The BP INTVL setting defines the amount of time between bump tests. The
minimum setting is 0 days and the maximum setting is 30 days. The factory
setting is 30 days.
NOTE:
If the BUMP DSP menu item is set to OFF, this menu item does not
appear.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the BP INTVL menu item.
Press and release the POWER ENTER button. The screen indicates the current
setting.
4GAS
MAINTENANCE
C--CHECK
30
MAINTENANCE
BP INTVL