150 • Using Maintenance Mode
Gas Tracer Operator’s Manual, CSA Version
2. Use the AIR
▲
or (SHIFT)
▼
button to display the desired setting.
Tip:
Press and hold the AIR
▲
or (SHIFT)
▼
button to rapidly scroll through the
settings.
3. Press and release the POWER ENTER button to save the setting and return to
the main menu.
Updating the Bump Test Reminder Setting
With BP RMNDR set to ON, the Gas Tracer will give an indication at start up if it
is due for bump testing. The type of indication will depend on the BP EXPRD
setting (see below).
With BP RMNDR set to OFF (factory setting), the Gas Tracer will give no
indication at start up if it is due for bump testing.
NOTE:
If the BUMP DSP menu item is set to OFF, this menu item does not
appear.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the BP RMNDR menu item.
Press and release the POWER ENTER button. The screen will indicate the
current setting.
2. Use the AIR
▲
or (SHIFT)
▼
button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and return to
the main menu.
Updating the Bump Test Expired Action Setting
This item defines what indication is given during start up when a bump test is
due and BP RMNDR is set to ON.
With BP EXPRD set to CONFIRM (factory setting), the Gas Tracer will give an
indication at start up if a bump test is due and require the user to press the RESET
SILENCE button to continue.
With BP EXPRD set to NOT USE, if the unit is due for a bump test the Gas Tracer
will give an indication at start up that the bump test is due and the unit cannot be
used until it is bump tested. With this setting, it is necessary to bump test the
instrument from the Maintenance Mode if a bump test is not performed before it
is due. See “Performing a Bump Test” on page 120 and “Calibrating and Bump
Testing the PPM Combustible Sensor” on page 125 for instructions.
With BP EXPRD set to NO EFFECT, no confirmation during startup or bump test
is required to use the Gas Tracer.
NOTE:
If the BUMP DSP menu item is set to OFF, this menu item does not
appear.
On
MAINTENANCE
BP RMNDR