Gas Tracer Operator’s Manual, CSA Version Calibrating and Bump Testing the PPM Com
-
Performing a PPM Bump Test
This section describes how to bump test the PPM combustible gas sensor.
NOTE:
The BUMP DSP menu item in Maintenance Mode is factory set to OFF.
The BUMP menu item will not appear unless BUMP DSP is set to ON.
See “Turning the Bump Test Function On or Off” on page 143 for
instructions.
NOTE:
The Gas Tracer uses parameters defined in the BUMP--SET menu item
of Maintenance Mode while performing a bump test. To view or update
these parameters, see “Updating the Bump Test Parameters” on
page 145.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the BUMP menu item.
2. Fill the gas bag with ppm calibration gas as described in “Filling the Gas Bag”
3. Install the hose and probe onto the Gas Tracer.
4. Press and release the POWER ENTER button. The bump test gas value will be
displayed.
5. Before proceeding, confirm that the bump test gas value is the same as the
concentration in the calibration cylinder. If it is not, adjust the bump test gas
value by entering the AUTO CAL menu item in the PPM CAL menu,
changing the value there, and reentering the BUMP menu.
6. Connect the humidifier from the gas bag to the rigid tube on the probe and
quickly press and release the POWER ENTER button. The unit will begin
counting down from 30.
MAINTENANCE
BUMP
500
ppm
MAINTENANCE
CH
4
BUMP 30