122 • Using Maintenance Mode
Gas Tracer Operator’s Manual, CSA Version
11. When AUTO CAL in the BUMP--SET menu item in Maintenance Mode is set
to OFF:
•
The instrument will display which channels passed or failed the bump
test with a P (pass) or an F (fail) to the right of the gas.
•
To view the bump test gas readings press the AIR
▲
or (SHIFT)
▼
button.
You can scroll between the bump test results and the bump test gas
readings with the AIR
▲
or (SHIFT)
▼
button.
•
To return to the BUMP screen in Calibration Mode, press POWER
ENTER.
If one or more sensors fail the bump test, see “Troubleshooting” on page 159 to
investigate the cause of the failure and replace the failed sensor or sensors if
necessary.
•