90 • Start Up, Bar Hole Mode
Gas Tracer Operator’s Manual, CSA Version
•
If the unit is due for a bump test and BP EXPRD is set to NOT USE, then
the following screen displays.
The Gas Tracer cannot be used until a bump test has been performed by
selecting BUMP in the Maintenance Mode menu. See “Performing a
Bump Test” on page 120 for bump testing instructions.
•
If a bump test is not due or BP EXPRD is set to NO EFFECT, then the
following screen appears for a few seconds indicating when the next
bump test is due.
NOTE:
If both CL RMNDR and BP RMNDR are turned off, a WARM UP
screen will display before the warm up sequence continues.
8. If ID DISP is turned on (see page 152), the User ID Screen displays for a few
seconds and then the Station ID Screen displays for a few seconds.
FAIL
B--LIMIT
2011
9
30
10d
NEXTBUMP
U Id
- - - - - - - -
S Id
- - - - - - - -