Gas Tracer Operator’s Manual, CSA Version
Bump Testing • 187
8. When AUTO CAL in the BUMP--SET menu item in Maintenance Mode is set
to OFF:
•
The instrument will display which channels passed or failed the bump
test with a P (pass) or an F (fail) to the right of the gas.
•
To view the bump test gas readings press the AIR
▲
or (SHIFT)
▼
button.
You can scroll between the bump test results and the bump test gas
readings with the AIR
▲
or (SHIFT)
▼
button.
•
To return to the BUMP screen in Calibration Mode, press POWER
ENTER.
See “Troubleshooting” on page 159 to investigate the cause of the failure and
replace the failed sensor or sensors if necessary.
•
Disconnect the tubing from the probe.