1-24
Chapter 1: Overview
System Limits
Note the following important ITS system limits.
IntelliVue Telemetry System Specifications
This section lists power, radio, and regulatory compliance specifications for the
IntelliVue Telemetry System.
Power Requirements
Table 1-11: Maximum ITS Capacities
ITS Device
Maximum Supported
Access Point Controllers
9 (includes one for redundancy)
1.4 GHz Standard Access Points
320 (routed ICN topology)
48 (non-routed ICN topology)
1.4 GHz Core Access Points
1.4 GHz Remote Antennas
320 (routed ICN topology)
48 (non-routed ICN topology)
640 (routed ICN topology)
96 (non-routed ICN topology)
2.4 GHz Access Points
2.4 GHz ITS does not support routed ICN topology.
48 (non-routed ICN topology)
1.4 GHz Patient-worn Transceivers
1024 (routed ICN topology)
128 (non-routed ICN topology)
2.4 GHz Patient-worn Transceivers
2.4 GHz ITS does not support routed ICN topology.
128 (non-routed ICN topology)
Table 1-12: ITS Component Power Requirements
Device/
Component
Input Voltage
(VAC/ VDC)
Manual
Switching
Required?
Input
Frequency
(Hz)
Dissipated
Power
(Max - Watts)
Standard or Core
Access Point
48 VDC Nominal
(44 - 52) VDC
287 mA
No
DC
< 13.8W
SYNC Unit
88-264 VAC
No
47-63
< 10W/18VA
POE with 1 AP
88-264 VAC
No
47-63
< 20W/50VA
POE with 12 AP
88-264 VAC
No
47-63
< 165W
Access Point
Controller
88-264 VAC
No
47-63
< 10W/40VA
Содержание IntelliVue Telemetry System Infrastructure
Страница 8: ...viii Contents ...
Страница 14: ...xiv About This Guide ...
Страница 100: ...3 22 Chapter 3 Installing and Configuring the ITS Figure 3 10 APC Filter Configuration Screen ...
Страница 102: ...3 24 Chapter 3 Installing and Configuring the ITS Figure 3 11 BOOTP DHCP Server Configuration Screen ...
Страница 162: ...5 12 Chapter 5 Troubleshooting and Testing ...
Страница 168: ...A 6 Chapter A Installing Multiple ITSs at a Single Hospital Site ...
Страница 180: ...Index 8 Index ...