ITS Access Point Controller Test and Inspection Procedures
IntelliVue Telemetry System Infrastructure Installation and Service Guide
5
-
3
ITS Access Point Controller Test and Inspection Procedures
Table 5-3 lists which tests you must perform when performing service events on
IntelliVue Telemetry System Access Point Controllers.
Table 5-3: ITS Access Point Controller Test and Inspection Requirements
When Performing this Service Event...
Complete These Required Test(s)
Installation
Visual, Power On, Performance
Preventive Maintenance
N/A
Any component repair or replacement
Power On, Performance
Hardware Upgrade
N/A
Software Upgrade
Power On, Performance, Revision Check
All other Service Events
Visual, Power On, Performance
Table 5-4: ITS Access Point Controller Test and Inspection Matrix
Test Block
Name
Test or Inspection to Perform
Expected Results
Information to
Record on
Service Record
APC Visual
Inspect all system components for obvious
damage. Visually inspect all system
components, cables, and connectors. Check
for signs of abrasion, wear, or other damage.
No visible damage
V:P or V:F
where:
P = Pass
F = Fail
APC Power On
With power connected to each active Network
device, observe that all lights visible on the
front panel are in proper status and that no
error conditions are shown. Following are
normal conditions for each type of device:
- Power LED - On GREEN
- Network Utilization LEDs -
See page 1-10.
- 100 BaseT LED - On GREEN
-LINK/Act - On primarily amber, flashes when
network activity is present
Devices power up into
expected status; no error
indications are shown.
PO:P or PO:F
where:
P = Pass
F = Fail
APC Performance
Does the APC appear in the APC web browser
management screens? Does data from a
wireless device associated with this APC
appear at the Information Center?
Expected answers are
“Yes”. If so, Performance
test passed.
P:P or P:F
where:
P = Pass
F = Fail
APC
Revision Check
Does the revision reported in the APC web
browser screen match the revision loaded?
Run the Philips Upgrade Wizard to check and
verify final configurations.
Expected answer is
“Yes”. If so, Revision
Check test passed.
RC:P or RC:F
where:
P = Pass
F = Fail
APC Safety
No safety test is required
S:NA
where:
NA=Not
Applicable
Содержание IntelliVue Telemetry System Infrastructure
Страница 8: ...viii Contents ...
Страница 14: ...xiv About This Guide ...
Страница 100: ...3 22 Chapter 3 Installing and Configuring the ITS Figure 3 10 APC Filter Configuration Screen ...
Страница 102: ...3 24 Chapter 3 Installing and Configuring the ITS Figure 3 11 BOOTP DHCP Server Configuration Screen ...
Страница 162: ...5 12 Chapter 5 Troubleshooting and Testing ...
Страница 168: ...A 6 Chapter A Installing Multiple ITSs at a Single Hospital Site ...