Starting Data Acquisition. 69
Equilibrate the System
Before running an analysis, the LC system must be equilibrated to achieve a stable chromatographic
baseline and to properly condition the LC column.
To equilibrate the system:
1.
In Chromera click
Run Time
then click
Manual Control
for the
Control Mode
.
2.
Make sure the SQ 300 MS in the
Operate
mode
.
Observe that in the
Status Panel
, the
MS Detector State
displays
Operate
. If not, Look at the
Manual Control
section of the
Run Time
screen.
3.
In the
Operate
row click
Apply.