26 .
Flexar SQ 300 MS User’s Guide
3.
Select
Auto Tune
from the
Tune
menu.
The
Auto Tune
dialog displays.
4.
In the
Auto Tune
dialog, select a tuning compound reference file by clicking on
Restore.
This displays the
Load Calibration Parameters
screen.
5.
For Positive ion, highlight
PKI tune mix positive up to 3000u
and simply double-click on it. If
calibration is only required up to 2000u, then the
PKI tune mix positive up to 2000u
file is
available to use.
After running the
PKI tune mix positive up to 3000u
you will run Auto Tune again using the
PKI Tune mix negative up to 3000u
.