Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector. 43
The
Mass
data evaluation screen appears.
14.
Left mouse click and drag a box around the reserpine protonated molecular ion cluster at
m/
z 609
and select
Zoom X-Y Axis
from the drop-down list.
15.
Move the cursor to the peak apex (it then turns into a hand), and right-click.
The peak table is displayed on the botton of the
Mass
data evaluation screen.