Creating Methods and Sequences. 57
10.
Select the most recent
pos_1k
tune file (with a date/time stamp), then click
OK
.
This example shows
pos_1k 29 Oct 2010
is selected as the most recent pos_1k tune file.
11.
In the
Global Variables
section, click on
Acquisition Function,
select
SIM
from the drop-down list,
and then press
Enter
. This will convert the
Tune
from a scanning acquisition to a SIM acquisition.
12.
To set the specific SIM mass, select
Low m/z,
type your value, and press
Enter
. This example
uses
m/
z
609.3
.
13.
Set the
Pulse Counting Time (µs)
to
300000
.
This is a 300 msec dwell time, which will give excellent signal to noise. It is possible to acquire
data with much shorter dwell times to achieve higher sampling rates, such as those required for
UHPLC separations or if the acquisition requires monitoring many SIM ions “simultaneously”
(current limit is 38).
14.
Set the
Drying Gas Flow Rate (L/min)
to
12
.
The optimal drying gas flow is dependent on a number of factors, such as the LC flow rate,
composition of the mobile phase and the temperature setting of the
Drying Gas
.
15.
Click on
Capillary Exit (Volts)
and from the drop-down list and select
Single Valve
. Enter a
value of
150
.
Содержание FLEXAR SQ 300 MS
Страница 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Страница 2: ......
Страница 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Страница 8: ...6 Flexar SQ 300 MS User s Guide ...
Страница 9: ...Starting ...
Страница 14: ...12 Flexar SQ 300 MS User s Guide ...
Страница 15: ...Starting Chromera ...
Страница 36: ...34 Flexar SQ 300 MS User s Guide ...
Страница 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Страница 55: ...Creating Methods and Sequences ...
Страница 68: ...66 Flexar SQ 300 MS User s Guide ...
Страница 69: ...Starting Data Acquisition ...
Страница 77: ...Analyze Results in Post Run ...
Страница 103: ...Analyze Results in Post Run 101 9 Select Print from the File menu in Microsoft Excel to print the table ...
Страница 104: ...102 Flexar SQ 300 MS User s Guide ...
Страница 105: ...Evaluating Mass Spectra ...
Страница 119: ......