50 .
Flexar SQ 300 MS User’s Guide
14.
Select
Tune
from the
View
menu.
Look at the
Capillary Exit (Volts)
value (the window below shows
Single=152
).
This means that the value of the Capillary Exit voltage at the cursor position, which was the
maximum of the curve, was equal to 152 volts. This is the Capillary Exit value that provides the
maximum intensity for the reserpine protonated molecular ion.
With a typical “bell shaped curve” as is demonstrated here, any value near the maximum will be
perfectly acceptable. So in this example, any value in between 150 to 160 volts will provide the
maximum signal intensity for reserpine.
15.
Close all
Snapshot
screens except the BIC and main screen.
Содержание FLEXAR SQ 300 MS
Страница 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Страница 2: ......
Страница 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Страница 8: ...6 Flexar SQ 300 MS User s Guide ...
Страница 9: ...Starting ...
Страница 14: ...12 Flexar SQ 300 MS User s Guide ...
Страница 15: ...Starting Chromera ...
Страница 36: ...34 Flexar SQ 300 MS User s Guide ...
Страница 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Страница 55: ...Creating Methods and Sequences ...
Страница 68: ...66 Flexar SQ 300 MS User s Guide ...
Страница 69: ...Starting Data Acquisition ...
Страница 77: ...Analyze Results in Post Run ...
Страница 103: ...Analyze Results in Post Run 101 9 Select Print from the File menu in Microsoft Excel to print the table ...
Страница 104: ...102 Flexar SQ 300 MS User s Guide ...
Страница 105: ...Evaluating Mass Spectra ...
Страница 119: ......