28 .
Flexar SQ 300 MS User’s Guide
10.
While viewing the
Mid-Range Mass
, you can adjust the source position by turning the indicated
controls for maximum sensitivity.
Locations of the Source Setting Adjustments.
Depth of probe adjustment for maximum
sensitivity.
NOTE:
After clicking the
Start Pumping
button, make sure you are getting flow
through the PEEK tubing. If there is no flow,
remove the tubing connection to the probe,
cut off about ¼-inch of tubing with a tubing
cutter, and reconnect the tubing to the probe.
If this fails to work, replace the tubing.