Starting. 11
7.
Check the vacuum status displayed in the
Status
screen.
When ready, the
Lamps (Power, Vacuum, and Ready)
should all be green.
The
Vacuum
lamp changes from flashing to steady light at this stage.
NOTE: The vacuum state light will take several hours to become ready and the system will not be able to load a
Tune file.
Checking the Position of the ESI Probe
1.
Observe the ESI probe through the inspection window and align to the center of the Capillary:
loosen the lock ring and move the manifold by turning the position adjustment screw. Tighten the
lock ring.
2.
The marks on the probe assembly can be used for quick positioning. Observe the needle through
the inspection window and adjust the tip if necessary.
3.
Loosen the lock ring.
4.
Turn the needle assembly adjustment screw until the needle tip protrudes about 0.5 mm from the
nozzle.
5.
Tighten the lock ring.
Содержание FLEXAR SQ 300 MS
Страница 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Страница 2: ......
Страница 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Страница 8: ...6 Flexar SQ 300 MS User s Guide ...
Страница 9: ...Starting ...
Страница 14: ...12 Flexar SQ 300 MS User s Guide ...
Страница 15: ...Starting Chromera ...
Страница 36: ...34 Flexar SQ 300 MS User s Guide ...
Страница 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Страница 55: ...Creating Methods and Sequences ...
Страница 68: ...66 Flexar SQ 300 MS User s Guide ...
Страница 69: ...Starting Data Acquisition ...
Страница 77: ...Analyze Results in Post Run ...
Страница 103: ...Analyze Results in Post Run 101 9 Select Print from the File menu in Microsoft Excel to print the table ...
Страница 104: ...102 Flexar SQ 300 MS User s Guide ...
Страница 105: ...Evaluating Mass Spectra ...
Страница 119: ......