Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector. 41
8.
To download to the SQ 300 MS all the changes made to the Tune, we must “Apply the Tune” by
selecting
Tune
from the
Apply
menu. This will start the syringe flowing at the specified rate.
9.
To start the data acquisition, click the green run button.
Acquire data for 1-2 minutes to allow enough time for the reserpine standard to be pumped all the
way to the ESI sprayer, so data can be collected on it.
10.
Take a “snapshot” of the data by selecting
Snapshot
from the
Collect
menu. A snapshot allows
preliminary processing of the data acquired up to the point in time the snapshot was executed.
The
TIC
and
Mass
data evaluation screens appear.
Содержание FLEXAR SQ 300 MS
Страница 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Страница 2: ......
Страница 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Страница 8: ...6 Flexar SQ 300 MS User s Guide ...
Страница 9: ...Starting ...
Страница 14: ...12 Flexar SQ 300 MS User s Guide ...
Страница 15: ...Starting Chromera ...
Страница 36: ...34 Flexar SQ 300 MS User s Guide ...
Страница 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Страница 55: ...Creating Methods and Sequences ...
Страница 68: ...66 Flexar SQ 300 MS User s Guide ...
Страница 69: ...Starting Data Acquisition ...
Страница 77: ...Analyze Results in Post Run ...
Страница 103: ...Analyze Results in Post Run 101 9 Select Print from the File menu in Microsoft Excel to print the table ...
Страница 104: ...102 Flexar SQ 300 MS User s Guide ...
Страница 105: ...Evaluating Mass Spectra ...
Страница 119: ......