54 .
Flexar SQ 300 MS User’s Guide
Creating an M S Method
The foundation for an optimal MS Method is starting with a mass calibrated Tune (with a date/time
stamp). These are created every time Auto Tune is completed and under normal operating conditions,
they can remain stable and usable for months
The following example shows how to create an MS method that will acquire both a Scan and a SIM for a
given period of time, alternating between the two acquisition modes. For SIM acquisitions, the correct
mass calibrated Tune to optimize for the acquisition will always be the
1k Tune
(with date/time stamp).
For Scan acquisitions on real world samples, selection of the correct mass calibrated Tune depends on
the
. However, if there is any doubt regarding whether the MS
is still properly calibrated, the easiest thing to do is to run a Check Tune to verify that the MS is operating
to specification. If Check Tune indicates that retuning is required, simply run Auto Tune to recalibrate the
system.
scan speed required to get enough sample points to properly profile the chromatographic
To create an MS method:
peak.
Typically, at least 10-15 data points (i.e., in this case, Scans) are required to properly profile a peak, and
the mass range required for the scan is another factor requiring consideration. The user should select
from the
1k
,
5k
and
10k
Tunes (with date/time stamps) that will best meet the scanning requirements.
1.
Click
Method
to open the
Chromera Method
screen.
2.
Select
Create/Edit MS Method/Tune
from the
File
menu.
The
SQ 300 MS driver
screen displays.
Содержание FLEXAR SQ 300 MS
Страница 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Страница 2: ......
Страница 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Страница 8: ...6 Flexar SQ 300 MS User s Guide ...
Страница 9: ...Starting ...
Страница 14: ...12 Flexar SQ 300 MS User s Guide ...
Страница 15: ...Starting Chromera ...
Страница 36: ...34 Flexar SQ 300 MS User s Guide ...
Страница 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Страница 55: ...Creating Methods and Sequences ...
Страница 68: ...66 Flexar SQ 300 MS User s Guide ...
Страница 69: ...Starting Data Acquisition ...
Страница 77: ...Analyze Results in Post Run ...
Страница 103: ...Analyze Results in Post Run 101 9 Select Print from the File menu in Microsoft Excel to print the table ...
Страница 104: ...102 Flexar SQ 300 MS User s Guide ...
Страница 105: ...Evaluating Mass Spectra ...
Страница 119: ......