30 .
Flexar SQ 300 MS User’s Guide
The
Status
light on the Auto tune page will turn
yellow
and displays
Status:
Executing auto tune
and the
Autotune Report
screen displays.
As Auto Tune runs in the positive ion mode, it creates the following three master tunes:
pos_1k,
pos_5k,
and
pos_10k
upon completion.
It also creates a copy of each master tune with a date and time stamp appended to the file name.
For example:
1.
Pos_1K_04_Oct_2010_16h21m
2.
Pos_5K_04_Oct_2010_16h21m
3.
Pos_10K_04_Oct_2010_16h21m
where:
Pos
means: Positive Ion Mode
Neg
means: Negative Ion Mode
1K
means: a scan rate of 1000 u/sec
5K
means: a scan rate of 5000 u/sec
10K
means: a scan rate of 10,000 u/sec
The time and date stamp tunes should be used to create any Tunes to be used in a method for data
acquisition. The date and time stamp allows the newest Tune to be distinguished from older ones. If
Auto Tune runs to completion and passes all criteria, the
Status
light turns
green
and
Status:
Auto tune Completed successfully
displays
.
Содержание FLEXAR SQ 300 MS
Страница 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Страница 2: ......
Страница 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Страница 8: ...6 Flexar SQ 300 MS User s Guide ...
Страница 9: ...Starting ...
Страница 14: ...12 Flexar SQ 300 MS User s Guide ...
Страница 15: ...Starting Chromera ...
Страница 36: ...34 Flexar SQ 300 MS User s Guide ...
Страница 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Страница 55: ...Creating Methods and Sequences ...
Страница 68: ...66 Flexar SQ 300 MS User s Guide ...
Страница 69: ...Starting Data Acquisition ...
Страница 77: ...Analyze Results in Post Run ...
Страница 103: ...Analyze Results in Post Run 101 9 Select Print from the File menu in Microsoft Excel to print the table ...
Страница 104: ...102 Flexar SQ 300 MS User s Guide ...
Страница 105: ...Evaluating Mass Spectra ...
Страница 119: ......