7
Maintenance
88
6924-en_V1.3_09/20
7.2.2
Calibrating particle size and particle velocity
Prerequisites
• The device’s zero point has been checked.
• The calibration menu is open.
Calibration interval
We recommend calibrating the aerosol sensor every week.
Administering calibration dust
1.
Connect a hose to the aerosol inlet of the aerosol sensor.
2.
Hold the hose in the container of
MonoDust 1500
.
3.
Gently squeeze the container several times to exchange some air be-
tween the container and the environment.
Mas-Q-Check firmware: Calibration menu with graphs containing signals