6
Operation and settings
6924-en_V1.3_09/20
65
MEASURE_PRESSURE | Duration[s] | register(dP0, dP1, dP2, dP3)
Function: Measurement of differential pressure.
Valid values for parameter 1
| Duration[s]
Meaning
Whole number
Duration of measurement in sec-
onds.
Valid values for parameter 2
| register(dP0, dP1, dP2, dP3)
Meaning
dp0
First value of differential pressure
measurement when inhaling.
dp1
Second value of differential pressure
measurement when inhaling.
dp2
Value of differential pressure mea-
surement when exhaling.
dp3
Value of differential pressure mea-
surement when testing valve func-
tion.
MEASURE_WEATHER | Duration[s] | type(SALT, OIL, BR)
Function: Measurement of temperature and relative humidity (possible only
when the respective sensors are installed).
Valid values for parameter 1
| Duration[s]
Meaning
Whole number
Duration of measurement in sec-
onds.
Valid values for parameter 2
| type(SALT, OIL, BR)
Meaning
SALT
Temperature and relative humidity of
saline aerosol.
OIL
Temperature and relative humidity of
oil aerosol.
BR
Temperature and relative humidity
when testing breath resistance.