6
Operation and settings
44
6924-en_V1.3_09/20
In the
script
menu, test routines can be started by selecting the respective
button. There are four standard testing routines:
Button
Description
Additional information
Efficiency Test
(salt)
Starts the testing routine for an efficiency test
with a saline aerosol based on EN 13274- 7.
Efficiency Test (oil)
Starts the testing routine for an efficiency test
with an oil aerosol based on EN 13274- 7.
Breath Resistance
Starts the testing routine for testing breath re-
sistance based on EN 149. The differential
pressure before and after the filter medium is
tested.
FFP 3 Valve Test
Starts the testing routine for testing the mask
valve based on EN 149.
Test routines can also be defined and programmed upon request. These
testing routines are saved under
further scripts
.