6
Operation and settings
6924-en_V1.3_09/20
73
1 Name of script
2 Last result for saline aerosol is
deleted
3 Dialog window: Operator input
4 Dialog window: Select a mask
5 Dialog window: Batch input
6 Prompt to close filter holder
7 Select aerosol: Salt aerosol
8 Set total flow rate
9 Waiting time: 5 seconds
10 Set aerosol flow rate
11 Waiting time: 60 seconds
12 Measurement over 20 seconds
without mask
13 Shut off aerosol flow
14 Waiting time: 3 seconds
15 Shut off total flow
16 Dialog window with graphic:
Load mask
17 Set total flow rate
18 Waiting time: 5 seconds
19 Set aerosol flow rate
20 Waiting time: 180 seconds
21 Measurement over 30 seconds
with mask
22 Shut off aerosol flow
23 Waiting time: 3 seconds
24 Shut off total flow
25 Save test report for saline
aerosol