Initial commissioning
6924-en_V1.3_09/20
31
Connection between
VKL 100
and sampling tube
1
2
3
Connection between VKL 100 and sampling tube
1 Sampling tube
2
VKL 100
3 Compressed air line (NW 13
mm), 6 – 8 bar, from the back
of the test rig
The exhaust fitting must be connected when depressurized to a filter and pu-
rification system belonging to the operator (system not included in delivery).
The exhaust fitting is intended to be used to connect a hose ID 28 mm.
WARNING
Escaping aerosols might be toxic
Aerosols can be hazardous to human health. If the aerosol outlet is not
connected to a suitable filter or purification system, aerosols can escape
into the environment and harm persons.
4
Before switching on: Verify that all aerosol outlets are connected to a
suitable filter and purification system.