7
Maintenance
86
6924-en_V1.3_09/20
Checking dilution factor
1.
Touch
expert user menu
.
ð
The
expert user menu
opens.
Firmware
Mas-Q-Check
:
Expert user menu
2.
Touch
sensor/calibration
.
ð
A menu showing sensor data opens.
Firmware
Mas-Q-Check
:
Sensor/calibration
3.
Check that the primary pressure is set to 2.0 bar on the dilution system
VKL 100
(default).
4.
Check the value of the parameter
dilution/calibration factor
:
The value has to be the same as the “Dilution factor W” shown on the
calibration certificate for the dilution system. The "dilution factor W" de-
pends on the set primary pressure of the dilution system.