6
Operation and settings
66
6924-en_V1.3_09/20
MEASURE_DISTRIBUTION | Duration[s] | source(UPSTREAM, DOWN-
STREAM)
Function: Measurement of number concentration and particle distribution.
Valid values for parameter 1
| Duration[s]
Meaning
Whole number
Duration of measurement in sec-
onds.
Valid values for parameter 2
| source(UPSTREAM, DOWN-
STREAM)
Meaning
UPSTREAM
Measurement of number concentra-
tion and particle distribution without
mask.
DOWNSTREAM
Measurement of number concentra-
tion and particle distribution with
mask.
MEASURE_FLAMEPHOTOMETER | Text | Unit | source(UPSTREAM,
DOWNSTREAM)
Function: Creation of a dialog window for entering the values measured by a
flame photometer (not included in the scope of delivery). The entered values
are used to calculate the penetration value (ratio of the two values).
Valid values for parameter 1
| Text
Meaning
Text
Text that explains the value to be
entered (e.g. “Enter measured
value”).
Valid values for parameter 2
| Unit
Meaning
Text
Text used to designate the unit of
the entered value (e.g. “mV”).
Valid values for parameter 3
| source(UPSTREAM, DOWN-
STREAM)
Meaning
UPSTREAM
Entered measured value without
mask
DOWNSTREAM
Entered measured value with mask